USER INTERFACE SCREENS
ICP Alarm Setup
This screen is accessed by selecting HOME -> ALARM SETUP.
Figure 11: ICP Alarm Setup Screen
Analysis Threshold Setup Screen
This screen is accessed by selecting HOME -> ANALYSIS SETUP.
Figure 12: Analysis Threshold Setup Screen
Data Download Screen
This screen is accessed by selecting HOME -> DATA DOWNLOAD.
Figure 13: Data Download Screen
Use this screen to select the HIGH and LOW alarm for the
mean ICP value.
The HIGH and LOW alarm can be turned ON/OFF with the
switch on the left (green indicates alarm is ON).
Alarm values must be within -50 to 150mmHg and the
HIGH alarm value must always be greater than the LOW
value.
Use this screen to select the HIGH and LOW levels used for
the Pressure Time Dose (PTD) and Time Above Threshold
calculations (refer to Figure 5 and 8).
The HIGH and LOW levels can be turned ON/OFF with the
switch on the left (green indicates that the alarm is ON).
The threshold values must be within -50 to 150mmHg and
the High value must always be greater than the Low value.
Use this screen to select which data to download to an
external USB memory drive.
Note that more than one item can be selected for
download. Refer to Data Download section for additional
information.
15