EN - ENGLISH
APPENDIX C: DIGITAL DATA STREAMING COMMUNICATION PROTOCOL
Mean ICP Trend Packet Description
Description
Start of Packet
Product ID
Device SN
Protocol Version
Packet Type
Sensor Serial Number
Date
Time
Mean ICP Value (1 through 6) (-/+)xxx.x
Alarm Status (1 through 6)
Data Marker (1 through 6)
Data Counter
Packet Checksum
End of Packet
System Status Packet Description
Description
Start of Packet
Product ID
Device SN
Protocol Version
Packet Type
Date
Time
Message Counter
Packet Checksum
End of Packet
String
*
CERELINK
CLKxxxxxxx
V01
ICPTR
SNxxxxxxx
YYYY/MM/DD
HH:mm:ss
[H, L, -]
xx
00001
0xFFFF (initial value)
\r\n
String
*
CERELINK
CLKxxxxxxx
V01
ICPNC
ICPNZ
or
FAULT
YYYY/MM/DD
HH:mm:ss
00001
0xFFFF (initial value)
\r\n
Size (bytes)
1
8
10
3
5
9
10
8
6 * 6
1 * 6
2 * 6
5
4
2
Size (bytes)
1
8
10
3
5
10
8
5
4
2
38
Details
-
-
-
-
Mean ICP Trend data
-
-
-
6 consecutive values
H = High alarm active
L = Low alarm active
(-) = No alarm active
xx = Data marker number
(-) = No marker
Increment with each message
Reset when 99999 is reached
CRC16
x^16 + x^12 + x^5 +1
-
Details
-
-
-
-
ICPNC = ICP sensor Not Connected
ICPNZ = ICP sensor Not Zeroed
or
FAULT = System is in a Fault State
-
-
Increment with each message
Reset when 99999 is reached
CRC16
x^16 + x^12 + x^5 +1
-