Hach TitraLab AT1000 Serie Manual Básico Del Usuario página 12

Ocultar thumbs Ver también para TitraLab AT1000 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49
The syringe outlet tube has a blue ring on it. If anti-diffusion tips are necessary, remove the pre-
installed outlet tube from the syringe and install the tube from the application kit with the pre-installed
anti-diffusion tip.
Push the outlet tubes into the tube holder slots so that they are correctly attached.
5.8 Install the sensor
5.8.1 Install the legacy adapter
Go to
Connect the sensor
on page 13 if no legacy adapter is included in the application kit.
1. Connect the measuring, reference and temperature sensors to the legacy adapter.
2. Make sure that the instrument display shows the Home screen. Connect the legacy adapter cable
to a sensor socket on the rear panel of the instrument.
12 English

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Titralab at1102Titralab at1112Titralab at1122Titralab at1222

Tabla de contenido