OBJ_DOKU-193013-005.fm Seite 30 Freitag, 12. Februar 2021 9:42 09
20132-183: APC probe 1500 A
Diameter
Length
Required diameter of endoscope
working channel
Type of beam
Max. electrical capacity
ACC unit setting
ICC unit setting
VIO unit setting
3.2
Probes with lateral argon beam, so-called Side-
Fire probes
20132-180: APC probe 2200 SW
Diameter
Length
Required diameter of endoscope
working channel
Type of beam
Max. electrical capacity
ACC unit setting
ICC unit setting
VIO unit setting
1.5 mm
1.5 m
> 1.8 mm
(A)xial Beam
4000 V
p
Do not use probe!
SPRAY COAG Mode,
Power limitation 50 W max.
Automatic
2.3 mm
2.2 m
> 2.5 mm
(S)ide Fire with (W)ide Beam
4300 V
p
KOAG Mode, SPRAY 2,
Max. intensity level 3
SPRAY COAG Mode,
Power limitation 90 W max.
Automatic
30