2. Immerse the product completely in a hydrogen peroxide
bath, ensuring that the product does not come into contact
with other components in the bath (immersion time: at least
5 minutes).
3. Screw the irrigation adapter (Art. No. 20132-203) onto the
APC probe. The irrigation adapter has to "click" into place.
4. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
5. Remove the syringe. Leave the irrigation adapter on the
product for the next processing steps.
6. Immerse the product completely in an ultrasonic cleaning
bath. Do not allow the product to come into contact with oth-
er components in the bath (immersion time in accordance
with the instructions provided by the manufacturer of the
cleaning agent, ultrasonic cleaning: 10 minutes).
7. Check the probe tip. If contamination is still visible on the in-
strument, repeat precleaning steps 1 to 4.
9.6

Manual cleaning and disinfection

Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 12.3.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.