I N T E L E C T ® M O B I L E 2 Q U I C K S TA RT G U I D E

General characteristics

Vacuum Range
Vacuum Modes
Pulsed Mode
Maximum Vacuum settings 2 to 10, +10mbar to
-10mbar per setting
Minimum Vacuum settings in 1 to 9, +10mbar to
-10mbar per setting
Hold Time in minimum & maximum vacuum settings,
0-20 seconds, in 1 second steps, +/-0�5 seconds

ULTRASOUND SPECIFICATIONS

Frequency
Duty Cycles
Pulse Repetition Rate
Pulse duration:
20
�0 to 600 mbar maximum +/-
Continuous or Pulsed
Continuous�
10 setting over vacuum range,
60 mbar per setting,
+10 mbar to 10 mbar per setting
1 MHz; 3 MHz
10%, 20%, 50%, Continuous
16, 48, or 100 Hz
1 -31�25 ms
Max (ON): 31�25 ms
Min (OFF): 5ms

OUTPUT POWER

5%
1cm
2
1MHz 3MHz 1MHz 3MHz 1MHz 3MHz 1MHz 3MHz
1
0.9
2W
1.8W
2.7W
3W
(*)
2W/
2W/
cm
cm
2
3W/
3W/
cm
cm
2
(*) An error of + 0.25 W can be measured with 1cm2
US applicator, pulse mode 100Hz at 10% or 20% Duty
Cycle.
2cm
5cm
2
2
1.5
1
2.5
2.7
3W
2W
5W
5.4W
4.5W
3W
7.5W
8.1W
2W/
2W/
2W/
2W/
cm
cm
cm
cm
2
2
2
2
2
3W/
3W/
3W/
3W/
cm
cm
cm
cm
2
2
2
2
2
10cm
2
6
6.8
12W
6.8W
18W
13.6W
2W/
1W/
cm
cm
2
2
3W/
2W/
cm
cm
2
2