Chattanooga Intelect Mobile 2 Guia De Inicio Rapido página 24

Ocultar thumbs Ver también para Intelect Mobile 2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
I N T E L E C T ® M O B I L E 2 Q U I C K S TA RT G U I D E
WAVEFORMS (CONTINUED)
VMS
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
VMS is a symmetrical biphasic waveform with a 100 µsec
interphase interval. Because the pulse is relatively short, the
waveform has a low skin load, making it suitable for applications
requiring high intensities, such as in muscle re-education
protocols.
Output Mode
Available on Channels
Treatment Time (Stim)
Treatment time (Combo)
Mode Selection
Output Intensity
Anti-Fatigue
Channel Mode
Cycle Time
Frequency
Phase Duration
Ramp
Set Intensity
Individual/both Channel Intensity Setting in Reciprocal and
IRMS
DC component
24
Electrodes
1, 2
1-60 min
1-30 min
CC or CV
0- 140 mA (CC) 0-140 V (CV)
Off or On
Single, Reciprocal, Co-Contract
Continuous or User Defined
1-200 pps
30-1,000 µsec
0-5 sec
Co-Contract modes
0-50mA
No
Microcurrent
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
Microcurrent is a monophasic waveform of very low intensity.
The literature reports beneficial effects of this waveform in the
treatment of wounds. The physiological working mechanism of this
effect is as yet not clearly understood. It is thought to stimulate
tissue healing by stimulating the 'current of injury', a current
which naturally occurs in healing tissue.
Output Mode
Available on channels
Treatment Time
Mode Selection
Output Intensity
Duty Cycle
Frequency
Polarity
IRMS
DC component
Electrodes
1, 2
1-60 Min
CC
0-1,000 μA
50%
0.1-1,000 pps
Positive, Negative, or Alternating
0- 1mA
No

Publicidad

Tabla de contenido
loading

Tabla de contenido