Unless otherwise specified, ultrasound
controls accuracy is:
Peak to Average Ratio:
Beam Nonuniformity Ratio
Beam Type
Treatment Time
GENERAL SYSTEM OPERATING AND
STORAGE TEMPERATURE
Operating Conditions
The device will meet its requirement under the
following conditions:
Temperature:
Relative Humidity:
Atmospheric Pressure:
Transport and Storage Conditions
The device will remain in proper condition under the
following conditions:
Temperature:
Relative Humidity:
Atmospheric Pressure:
Time required for the Intelect Mobile 2 to warm from
the minimum storage temperature between uses until
the Intelect Mobile 2 is ready for its INTENDED USE
when the ambient temperature is 20 °C: 5h
± 20 %�
1:1, at 50% Duty
Cycle
4:1, at 20% Duty Cycle
9:1, at 10% Duty Cycle
<5:1
Collimating
1 to 30 min
5°C to 40°C
15% to 90%
70kPa to 106kPa
-20°C to 60°C
10% to 90%
50kPa to 106kPa
Time required for the Intelect Mobile 2 to cool from
the maximum storage temperature between uses until
the Intelect Mobile 2 is ready for its INTENDED USE
when the
ambient temperature is 20 °C: 5h
IPXX Rating for Unit
Rated to IP21
IP2* Protection against fingers or other object not
greater than 80mm in length and 12mm in diameter
*1 Protection from vertically dripping water
IPXX Rating for US applicator
Rated to IPX7
IPX7 Protection from immersed in water (up to 1m
depth)
RED
RF transmitter/receiver characteristics:
- Frequency Band transmission:
- Modulation type:
- Data rate:
up to 2Mbps 500kHz deviation
- Effective radiated power:
E N
2400–2483�5 MHz
GFSK
at 2Mbps
+6dBm
21