Waveforms - Chattanooga Intelect Mobile 2 Guia De Inicio Rapido

Ocultar thumbs Ver también para Intelect Mobile 2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
I N T E L E C T ® M O B I L E 2 Q U I C K S TA RT G U I D E

WAVEFORMS

Advice on size and type of electrodes to be used is
given in device GUI treatment guidelines�
CC: Constant Current
IFC (Interferential) Traditional (4 Pole)
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
Interferential Current is a medium frequency waveform. Current is
distributed through two channels (four electrodes). The currents
cross each other in the body at the area requiring treatment. The
two currents interfere with each other at this crossing point,
resulting in a modulation of the intensity (the current intensity
increases and decreases at a regular frequency).
Output Mode
Available on Channel
Treatment Time
Mode Selection
Output Intensity
Beat Frequency
Carrier Frequency
Cycle Time
Sweep Time
Sweep Low Beat Frequency
Sweep High Beat Frequency
Scan Percentage
IRMS
DC component
22
CV: Constant Voltage
Electrodes
1&2,
1-60 Minutes
CC
0-100 mA (CC)
1-200 Hz
2000-10,000 Hz
Continuous or User Defined
14 sec
1-199 Hz
2-200 Hz
Static, 40%, 100%, Manual
0-78mA
No
TENS- Asymmetrical Biphasic
Advice on size and type of electrodes to be used is given in device GUI
"treatment guidelines" feature
The Asymmetrical Biphasic waveform has a short pulse duration.
It is capable of strong stimulation of the nerve fibers in the skin
as well as of muscle tissue. This waveform is often used in TENS
devices.Because of its short pulse, the patient typically tolerates
the current well, even at relatively high intensities.
Output Mode
Output Intensity
Available on Channel
Treatment Time (Stim)
Treatment Time (Combo)
Mode Selection (Stim)
Mode Selection (Combo)
Amplitude Modulation
Burst Frequency
Cycle Time
Frequency
FrequencySweep
Phase Duration
Sweep time
Sweep Low Frequency
Sweep High Frequency
IRMS
DC component
Electrodes
0-140 mA (CC) 0-140 V (CV)
1,2
1-60 minutes
1-30 minutes
CC or CV
CV
0% (off) to 100% on 10% steps
0-10 bps
Continuous or User Defined
1-200 pps
On/Off
30-400 µsec
14 sec
1-199 pps
2-200 pps
0-50mA
No