Inspection

Inspect the device on a continual basis for unacceptable deterioration such as (but not limited
to) corrosion, discoloration, pitting, cracked seals, or abnormal noises. If a problem is observed or
suspected, the device should be returned for service.

Electrical Safety Testing

Testing of ground protective earthing from the display is not necessary. Protective earthing is
assured by the certified external power supply.

Storage

Never store the device in a non-ventilated, humid environment. This can damage the delicate
electronics in the device.

Expected Service Life

The expected service life of the device is four years.

Disposal

This product contains electrical waste or electronic equipment. It must not be disposed
of as unsorted municipal waste and must be collected separately in accordance with
applicable national or institutional related policies relating to obsolete electronic
equipment.
Dispose of any system accessories according to normal institutional practice relating to
potentially contaminated items.