Before returning your display for service, consult the troubleshooting list below:
Problem
No picture
Abnormal picture
OSD Error message
Current Status
Power LED on
Power LED off
Power LED blinking
Converters not
powering on
Oversized,
undersized or
missing video
Resolution not
supported
Remedy
Using the OSD Menu, adjust the brightness
and contrast to maximum or reset them to
their default settings.
Check the power supply functionality.
Ensure the power switch at the front and rear
of the display are set to ON.
Check if the AC power cord is properly
connected to the AC adapter and outlet.
Check that the power supply is fully
connected and functioning properly.
Display is in sleep mode. Connect ACTIVE
sources or change INPUT to wake the display.
Check that the converter power setting is ON
in the OSD menu.
Turn the hard power switch off and on again.
Adjust settings in Scale Mode.
Wait a few seconds after initial sync of video
signals, or power cycle the display.
Ensure the video source has a compatible
output resolution.