107
Anhang
Recommended separation distances between portable and mobile RF communications equipment
The UNIDRIVE
®
S III ENT model 40 7016 20-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters [m] can be estimated using the equa-
tion applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts [W] according to the transmitter
manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
Appendix
Table 206
and the UNIDRIVE
®
S III ENT model 40 7016 20-1
Separation distance d [m]
according to frequency of transmitter
150 kHz to < 80 MHz
80 MHz to < 800 MHz
D
0.12
0.38
1.2
3.8
12
Anexo
800 MHz to 2.5 GHz
D
D
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23