OBJ_DOKU-194305-003.fm Seite 29 Montag, 5. Oktober 2020 10:12 10
7
Cleaning, disinfection, sterilization
7.1
Safety instructions for processing
The product must be cleaned and disinfected while it is disman-
tled and the jaws are open.
The product may also be sterilized while assembled (not permit-
ted in the USA!). The jaws must be open.
The instrument is also suitable for cleaning in an ultrasonic bath.
It is absolutely essential that the instrument parts lie on a non-
slip surface. The instrument parts must not come into contact
with other instruments, instrument parts, or the tank.
Process the product immediately after use.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
Do not sterilize with gamma radiation.
29