2 Safety

2 Safety
When operating the device, comply with the following safety instructions as well as those contained in
subsequent chapters. Use the device only for the purpose specified in these instructions. Failure to comply
with the safety instructions can result in injury and/or damage to the device.
Qualified personnel
As with the preceding instrument reprocessing, only
n
steam sterilizer.
The operator must ensure that the users have been trained in the operation and safe handling of the device.
n
The operator must ensure that users are regularly trained in the operation and safe handling of the device.
n
Setup, installation and commissioning
Check the device after unpacking for any damage suffered during transport.
n
The device should only be setup, installed and commissioned by MELAG authorized persons.
n
The connections for electrical provision and water supply and discharge must be setup by trained personnel.
n
Using the optional electronic leak detector (water stop) minimizes the risk of water damage.
n
The device is not suitable for operation in explosive atmospheres.
n
Install and operate the device in a frost-free environment.
n
The device is conceived for use outside the patient area. The device should be located a minimum of 1.5 m radius
n
away from the treatment area.
Power cable and power plug
Only the power cable included in the scope of delivery may be connected to the device.
n
The power cable may not be replaced by a cable determined to be insufficient.
n
Comply with all legal requirements and locally-specified connection conditions.
n
Never operate the device if the plug or power cable are damaged.
n
The power cable or plug should only be replaced by
n
Never damage or alter the power plug or cable.
n
Never unplug by pulling on the power cable. Always take a grip on the plug.
n
Ensure that the power cable does not become jammed in.
n
Never lead the cable along a source of heat.
n
Never fix the power cable with sharp objects.
n
The mains socket must be freely accessible after installation so that the device can be disconnected from the electri-
n
cal mains at any time if necessary by pulling the mains plug.
Spring safety valve
The free movement of the valve rod must be guaranteed. For example, the spring safety valve must not be taped or
n
blocked. Install the device in such a way that the proper functioning of the spring safety valve is guaranteed.
Reprocessing and sterilization
Follow the manufacturer's instructions of your textile articles and instruments regarding their reprocessing and sterili-
n
zation.
Comply with the relevant standards and directives applicable to the reprocessing and sterilization of textiles and in-
n
struments (in Germany e.g. from the
6
competent personnel
}
 
authorized
}
 
RKI
and
DGSV).
}
}
 
 
should undertake sterilization using this
technicians.