Manual program abort

You can abort the program at any time. If you abort the program before the end of the sterilization phase, the load is not
sterile.
1.
Press ABORT to abort a program.
2.
Confirm the security query with YES.
Ê
The load is not sterile.
Ê
Aborting the program may take a few minutes as steam and
condensate are removed from the chamber.
CAUTION
When the door is opened after a program abort, hot steam may escape and hot water may be present
in the sterilization chamber.
This could result in scalding.
Never touch the load, the sterilization chamber or the door with unprotected hands. The components are
n
hot.
Press OPEN DOOR to remove the load.