Event
Possible cause
11011
Several printers are directly connected to the
device
11012
The paper of the printer will be used up
soon.
11100
The log output was aborted due to a
connection error.
What you can do
Please connect only one printer to the
device.
Please restart the printer.
First start the device and then the printer.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Please have a new roll ready.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Please check the connection of the device to
the practice network via the network
interface on the back of the device.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
14 Malfunctions