15 Technical data
15 Technical data
Device type
Device dimensions (H x W x D)
Empty weight
Operating weight
Sterilization chamber
Chamber diameter
Chamber depth
Chamber volume/steam generator
Electrical connection
Power supply
Max. power consumption in operation
Building fuse protection
Length of the power cable
Overvoltage category
Degree of contamination
(in accordance with EN 61010)
Ambient conditions
Installation location
Noise emission LP(a)
in 1 m distance
Heat emission per hour (with maximum
load)
Ambient temperature
Relative humidity
Degree of protection
(in accordance with IEC 60529)
Transport and storage conditions
Max. altitude
Feed water supply
Max. water consumption
Average water consumption
Water temperature
Min. static water pressure
Max. static water pressure
Water quality
Cold water supply (for connection of the water treatment unit)
Min. static water pressure
Max. static water pressure
Water quality
Wastewater connection
Max. throughflow volume
Max. water temperature
Working and operating pressures
68
Vacuclave 550
65.0 x 63.6 x 71.5 cm
98 kg
127 kg
380 mm
450 mm
53 l/12.5 l
220-230 V 50/60 Hz
max. voltage range 198-253 V
3400 W (15 A operation)
2700 W (13 A operation)
16 A, RCD with rated residual current = 30 mA (15 A operation)
13 A, RCD with rated residual current = 30 mA (13 A operation)
2 m
transient overvoltages up to the values of overvoltage category II
2
interior of a building
64 dB(A)
2.25 kWh
5-40 °C (ideal range 16-26 °C)
max. 80% at temperatures of up to 31 °C, max. 50% at 40 °C (decreasing in
linear fashion in-between)
IP20
Temperature: -18 to +50 °C, air humidity: < 80%
3000 m
5.5 l/cycle
2 l/cycle
1 to 35 °C
1 bar
10 bar
distilled or demineralized water in accordance with EN 13060, Appendix C
2 bar
10 bar
potable water
0.5 l/min
90 °C for 30 s, max. 98 °C for 1 s