19 - Preparing the ABPM 7100
Setting the time / date
The ABPM 7100 has an internal buffer battery allowing the time to continue even if the batteries have been removed.
Nevertheless the time and the date should be checked before every measurement series.
The time and date can be set automatically with the patient management software.
Alternatively the time and date can be set manually. Press and hold the START button and then press the EVENT
buttons to enter the Set Time mode. Use the START button to select the appropriate item and use the EVENT button
to jump to the next display item.
Clearing the memory
The device memory must be cleared before every measurement series, i.e. blood pressure data form the previous
patient must not remain in the memory.
If there are existing data, the memory can be cleared with the delete function of the analysis software.
Alternatively the data can be cleared manually. Press and hold the START button for a minimum of 5 seconds until
"cLr" is displayed. Within the next 5 seconds press and hold the EVENT button for at least 2 seconds to confirm the
deletion of the stored measurements. The device emits a single beep to indicate that the memory is cleared.
Transferring patient data
The ABPM 7100 must be prepared by transferring patient data (ID) with the help of the patient management software,
so that correct data allocation is possible when it is read out after measurement. Please refer to the respective patient
management software manual for how to transfer patient data (ID) to the ABPM 7100.
Setting measurement logs
In the patient management software you can optionally choose between eleven (1-11) logs. A log serves to set the
measurement intervals. As soon as you have conducted a measurement, the log can only be changed once you
have fully deleted all data.
Manual log settings
For manual log setting, press and hold the DAY/NIGHT button while simultaneously pressing the EVENT button. Use
the START button to change the log and confirm with the EVENT button.
Measurements
Log
Day-Time
Night-Time
per hour
1
08:00
23:59
4
00:00
07:59
2
2
08:00
22:59
4
23:00
07:59
1
3
07:00
21:59
4
22:00
06:59
2
4
08:00
23:59
4
00:00
07:59
2
5
18:00
09:59
4
10:00
17:59
2
6
07:00
23:59
4
00:00
06:59
2
7
06:00
22:59
4
23:00
05:59
2
8
07:00
08:59
6
09:00
23:59
4
00:00
06:59
2
9
09:00
08:59
30
10
08:00
07:59
30
11
08:00
23:59
4
00:00
07:59
2
Setting the logs via software
To set the logs via software please refer to the respective patient management software manual.
•
Note
Logs 1, 2 and 11 are set by default but can be changed via the patient management
software.
•
Log 5 is suitable for nighttime activities (night shift).
•
Log 9 is designated as "Schellong-Test".
•
Log 11 is only available to upgraded ABPM 7100 systems in connection with HMS from
version 5.0. Blood pressure measurement intervals and the 24h PWA can be set
separately here. Please contact Welch Allyn for further information.
Preparing the ABPM 7100 - 20