button" to enter next hour; Press "left button", it will show "-" in the position, then press "Down button"
to enter last hour.
D. The Review Interface 2 shown based in Review Interface 1, the stored SpO2 value and PR value in
each second can be observed here, the underside date from left to right marks time, SpO2 value, PR
value. When the stored data exceeds the upper and lower limit set by user, the relevant value will
turn green.
E. Press "up button" to exit the review Interface, return to the measure interface.
5.4 Clock interface
In the measure interface, press the "right button" can enter the clock interface of Figure 8.Press the
"right button" again can return to the measure interface.
Figure 8. Clock interface

5.5 Menu operations:

In the measure interface,press the "menu button "can enter the menu of Figure9.Users can adjust the
setting through the main menu, such as the sound prompt, record, clock., system, etc. can be set,
methods are as followings:
5.5.1 Sound prompt setting
Under main menu, press the "up button" or "down button" to select "Prompt", then press the "up button"
or "down button" to enter its setting interface shown in Figure 10.
Press the "up button" or "down button" to select the option to be adjusted, then press the "left button" or
"right button" to change the value.
"SpO2 HI(%)": upper limit prompt for SpO2 over-limit
"SpO2 LO(%)": lower limit prompt for SpO2 over-limit
"PR HI(bpm)": upper limit prompt for PR over-limit
"PR LO(bpm)": lower limit prompt for PR over-limit
"Prompt Sound": prompt for over-limit, "off": close, "on": open.
"Pulse Sound": PR sound, "off": close, "on": open.
Lower limit can not exceed the upper limit, and the upper limit can not be lower than the lower limit when
adjusting the values. SpO2 range: 0 % ~ 100 %, PR range: 0 ~ 254 bpm
The values displayed in Figure 10 are the initial values of over-limit prompt.
After setting, press the "menu button" to exit the Prompt Settings Menu interface, and return to "Main
Menu" interface.

5.5.2 Data storage

Under the main menu, press the "up button" or "down button" to select "Record", then press the "up
button" or "down button" to enter the "Record Menu" interface as shown in Figure 11.
Press the "up button" or "down button" to select the option to be adjusted, then press the "left button" or
"right button" to change the value.
It indicates that the device is storing when the red dot "R●" in measurement interface flickers "Mode":
record mode selection, including: "Auto" and "Manual" mode. Under "Manual" mode, select to turn on /
off memory by "Record".
ENGLISH
Figure 9. Main Menu
Figure 10 Setting for sound prompt