When the device is in storing , it can't be turned off.
6 MAINTAIN, TRANSPORT AND STORAGE

6.1 Cleaning and disinfection

The device must be turned off before cleaning, and it should not be immersed into liquid.
Please take out the internal battery before cleaning, do not immerse it into liquid.
Use 75% alcohol to wipe the device enclosure, and use liquid soap or isopropanol to wipe the watch-
band for disinfection, nature dry or clean it with clean and soft cloth. Do not spray any liquid on the
device directly, and avoid liquid penetrating into the device.
6.2 Maintenance
A. Check the main unit and all accessories periodically to make sure that there is no visible damage that
may affect patient's safety and monitoring performance. It is recommended that the device should be
inspected weekly at least. When there is obvious damage, stop using it.
B. Please clean and disinfect the device before/after using it according to the User Manual (6.1).
C. Please replace the batteries in time when low-battery appears.
D. Please take out the batteries if the device is not used for a long time.
E. The device need not to be calibrated during maintenance.

6.3 Transport and Storage

A. The packed device can be transported by ordinary conveyance or according to transport contract.
During transportation, avoid strong shock, vibration and splashing with rain or snow, and it can not
be transported mixed with toxic, harmful, corrosive material.
B. The packed device should be stored in room with no corrosive gases and good ventilation. Temper-
ature: -40°C~+60°C; Relative humidity: ≤95%.
7 TROUBLESHOOTING
Trouble
The values can not be
displayed normally or
stably.
The device can not be
turned on
The display disappears
suddenly.
The data can not be
stored.
Possible Reason
1) The finger is not properly inserted.
2) The finger is shaking or the patient is
moving.
3) The device is not used in environment
required by the manual.
4) The device works abnormally.
1) The battery is drained away or almost
drained away.
2) The battery is installed incorrectly.
3) The device's malfunction.
1) The device enters into the energy saving
mode.
2) Low battery.
3) The device works abnormally.
1) The device is not operated according to
the manual.
2) The device works abnormally.
ENGLISH
Solution
1) Please insert the finger properly and
measure again.
2) Let the patient keep calm.
3) Please use the device in normal envi-
ronment.
4) Please contact the after-sales.
1) Please change batteries.
2) Please Install the battery again.
3) Please contact the local service center.
1) Normal.
2) Please charge the battery.
3) Please contact the after-sales.
1) Please operate the device according to
the manual.
2) Please contact the after-sales.