Table 2: Possible Adverse Effects
• Total occlusion of the treated artery
• Arterial dissection or perforation
• Arterial spasm
• Pseudo-aneurysm
• Re-stenosis of the dilated artery
• Embolism
• Thrombus
• Retained device components
• Hemorrhage or hematoma
• Arteriovenous fistula
VIII. MATERIALS REQUIRED FOR USE WITH THE
ANGIOSCULPT® CATHETER:
WARNING: Use single use items only. Do not resterilize
or reuse.
• Femoral introducer sheath and/or guiding catheter (GC):
• 2.0, 2.5, 3.0, or 3.5 mm diameter × 20 and 40mm length
balloon catheter: ≥ 5F introducer sheath and/or ≥ 6F GC
• 2.0, 2.5, 3.0, or 3.5 mm diameter × 100mm length
balloon catheter and 4.0, 5.0, or 6.0 mm diameter
balloon catheter: ≥ 6F introducer sheath and/or ≥ 7F GC
• Hemostatic valve
• Radiographic contrast medium diluted ~1:1 with
normal saline
• Sterile heparinized normal saline
• 10-cc and 20-cc syringes for flushing and balloon prep
• Inflation device (indeflator)
• Guide wire:
• 2.0, 2.5, 3.0, 3.5 mm diameter and 4.0, 5.0 and
6.0 mm × 100 mm balloon sizes - 0.014"
• 4.0, 5.0, or 6.0 mm × 20 mm and 40 mm balloon
sizes - 0.018" only
• Guide wire introducer
• Guide wire torque device
• Manifold (for pressure monitoring and contrast injection),
extension pressure tubing
IX. INSTRUCTIONS FOR USE
Prior to use of the AngioSculpt, examine carefully for damage
and device integrity. Do not use if the catheter has bends, kinks,
missing components or other damage. Do not use if inner
package is open or damaged.
1. Premedicate patients with anti-coagulants, anti-platelet
agents and vasodilators according to institutional
protocol for PTA procedures.
2. Perform peripheral angiogram in the view
best demonstrating the target lesion prior to
device deployment.
3. Utilizing standard fluoroscopic technique, position
an appropriately sized guide wire of choice beyond
the target lesion (use an exchange length 300 cm
guide wire with the 137 cm and 155 cm long
Page 2 of 44
AngioSculpt catheters).
4. Using sterile technique, remove an appropriately sized
AngioSculpt catheter from the sterile package and
place on the sterile field.
• For the 2.0, 2.5, 3.0, or 3.5 mm balloon catheter,
size the AngioSculpt ≤ 1.0 x the reference vessel
diameter (RVD).
• For the 4.0, 5.0, or 6.0 mm balloon catheter, size the
AngioSculpt a minimum of 0.5 mm LESS than the
reference vessel diameter (RVD).
5. Remove the protective tubing from the balloon
(100 mm length balloons only).
6. Inspect the balloon catheter to ensure that all
components are intact.
7. Flush the guide wire lumen by connecting the proximal
hub to a 10-cc syringe and injecting heparinized saline
into the proximal guide wire lumen until droplets
emerge from the distal end.
8. Attach a stopcock to the catheter's balloon inflation port.
9. Attach a 20-cc syringe filled with 2-3 cc of 1:1 mixture of
radiographic contrast and normal saline to the stopcock.
10. Open the stopcock to the syringe, aspirate/remove
air from the catheter balloon lumen using the 20-cc
syringe filled with 2-3 cc of radiographic contrast and
leave on vacuum for 30 seconds.
11. Close the stopcock to the catheter balloon inflation port
and remove the syringe.
12. Attach inflation device (indeflator), filled with 1:1
mixture of radiographic contrast and normal saline, to
the stopcock by creating a meniscus. Avoid introducing
air bubbles into the catheter balloon lumen.
13. Open the stopcock to the inflation device and aspirate
using the inflation device, locking in vacuum.
NOTE: All air must be removed from the balloon and
displaced with contrast medium prior to inserting
into the body (repeat steps 8-12, if necessary).
14. Advance the AngioSculpt catheter over the guide wire
(through either a previously placed and appropriately
sized hemostatic introducer sheath or guiding catheter)
and position at the target lesion utilizing standard
fluoroscopic technique.
NOTE: When backloading the catheter onto the
guide wire, the catheter should be supported,
ensuring that the guide wire does not come in
contact with the balloon. Do not advance or
retract the AngioSculpt catheter over the floppy
portion of the guide wire. Do not advance or
retract the catheter unless the balloon is fully
deflated under vacuum. If resistance is met during
manipulation, determine the cause of the resistance
before proceeding.
15. Inflate the AngioSculpt balloon per the following
recommended protocol:
• increase the inflation pressure by 2 atmospheres
every 10-15 seconds until full balloon inflation
is achieved.
• do not exceed the rated burst pressure (RBP) printed
on the package label.
16. Apply negative pressure to the inflation device and
confirm that the balloon is fully deflated prior to
removing the AngioSculpt catheter.
17. Remove the AngioSculpt catheter.
NOTE: Do not rotate the catheter shaft in excess
of 180 degrees when the tip is constrained. Do not
rotate the catheter luer hub in excess of five (5)
turns during use. Catheter manipulation, including
advancement and retraction, should be performed
by grasping the shaft.
18. Perform peripheral angiogram (in the same view(s)
as step 2) of the target lesion following completion of
AngioSculpt treatment.
19. Inspect all components to ensure that the catheter is
intact. Follow institutional procedures for disposal of
biohazards. If device malfunction occurs or any defects
are noted on inspection, flush the guide wire lumen
and clean the outer surface of the catheter with saline,
store the catheter in a sealed plastic bag, and contact
Spectranetics® for further instructions.
20. Complete any additional interventions as clinically
indicated (e.g. stent placement).
21. Remove the guide wire and perform peripheral
angiography (in the same view(s) as step 2) of the target
lesion following completion of all interventions.
22. Remove all catheters and manage the arterial access site
according to institutional protocol.
23. Continue treatment with anti-coagulants, anti-platelet
agents and vasodilators according to institutional
protocol for PTA procedures.
X.
REFERENCES
The physician should consult recent literature on current
medical practice regarding balloon dilatation and
PTA procedures.
XI. DISCLAIMER OF WARRANTY AND LIMITATION
OF REMEDY
There is no express or implied warranty, including any
implied warranty of merchantability or fitness for a
particular purpose, on the Spectranetics® product(s)
described in this publication. Under no circumstances
shall Spectranetics® be liable for any direct, incidental
or consequential damages other than as expressly
provided by specific law. No person has the authority to
bind Spectranetics® to any representation or warranty
except as specifically set forth herein.
Descriptions or specifications in Spectranetics® printed
matter, including this publication, are meant solely to
generally describe the product at the time of manufacture
and do not constitute any express warranties.
Spectranetics® assumes no liability with respect to
instruments reused, reprocessed or resterilized.