5.1.6. Disconnect from Wi-Fi network:
• In the Network menu, press the currently selected Wi-Fi network, then press Disconnect.
5.1.7. Clear all network data from the displaying unit:
• In the Network menu, press Clear all data.
• Press OK.
5.2. PACS and Worklist setup
You can transfer patient data between the displaying unit and external servers. Before procedure
you can retrieve patient data from the Worklist server, including information about an upcoming
procedure. After procedure, the images and videos created during procedure can be exported in
DICOM format from the Archive to a PACS server. Before setting up server connections, ensure
that the displaying unit is connected to a Wi-Fi or LAN network (see section 5.1.).
5.2.1. Set up the displaying unit for server connection:
• Press the Settings tab, then press Setup.
• Press DICOM setup 22 .
• Press the input field next to Station name 23 and enter the device name.
The default name is AmbuMon
Note: The maximum length of the station name is 16 characters.
• Press Use serial number or Use custom name next to Station AE title 24 . If you selected
Use custom name, press the input field and enter the name.
5.2.2. Set up connection to PACS server:
• Press Add new 25 below PACS servers.
• Press the input field next to PACS name 26 and enter the name you want to use for the
PACS connection.
• Press the input fields next to PACS AE title, Host name and Port number 27 and enter the
required information in each field.
• Press the required setting next to TLS 28 .
Note: If you enable TLS, you need to import the required TLS certificate from a server or USB flash drive
(see instructions in section 5.2.3.).
• Press Test connection 29 to verify that the information has been entered correctly and the
server connection can be established.
• Press Create 30 to save the server connection setup.
18
24
22
25
28
30
26
27
29
23