5.2 Processing limits
Nebuliser, accessories, LC
tracheostomy adapter, PARI
tracheostomy mask, PARI
Tracheo mask adapter,
disinfection
Nebuliser, accessories, LC
tracheostomy adapter, PARI
tracheostomy mask and PARI
Tracheo mask adapter,
sterilisation
Connection tubing
5.3 Nebuliser
Separated parts for processing
CAUTION
Risk of infection due to cross-contamination in the case
of a change in patients
If a product is used for more than one patient, there is a risk
that germs may be transmitted from one patient to the next.
• Clean, disinfect and sterilise all separated parts before
every patient change.
• Replace the connection tubing or carry out mechanical
cleaning and disinfection of the connection tubing
[see: Connection tubing, page 38].
All components of a PARI nebuliser and the PARI accessories
used can be cleaned, disinfected and sterilised according to
the procedures described below.
The connection tubing must be treated separately.
– 31 –
300 processing cycles,
max. 1 year
100 processing cycles,
max. 1 year
50 reprocessing cycles,
max. 1 year