Descargar Imprimir esta página

Pari LC SPRINT Tracheo Manual De Instrucciones página 82

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 222
5.4 Connection tubing
Mechanical cleaning and disinfecting
EQUIPMENT:
The method has been validated in Europe using:
– Alkaline cleaning agent: Dr. Weigert neodisher
forte
– Neutralising agent: Dr. Weigert neodisher Z
– Cleaning and disinfection device: RDG G7836 CD (Miele)
(conforming to DIN EN ISO 15883)
– Special baskets for Miele instrument dishwasher
– Compressed air source for drying
PROCEDURE
Vario TD program or comparable valid programs
DRYING
Dry the connection tubing as described in the section on this
topic.
Drying
• Connect the connection tubing to a compressed air source
(compressor or central medical gas supply system).
• Switch the compressed air source on.
• Leave the compressed air source running until all the mois-
ture in the tube has been removed.
– 38 –
®
MediClean

Publicidad

loading

Este manual también es adecuado para:

023