Chemical disinfection
To ensure safety when handling chemicals, follow the instruc-
tions for use of the disinfectant.
EQUIPMENT:
The method has been validated in Europe using:
– Aldehyde-containing disinfectant: Bode Korsolex
(concentration: 4%)
Active agent basis: Aldehyde donor, aldehyde
– Application time: 30 minutes
PROCEDURE:
CAUTION
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Observe the mixing proportions indicated and the spe-
cified treatment time.
• Ensure that all components are completely submerged in
the solution for the whole of the treatment time. There
must not be any air pockets or bubbles.
• Disinfect the individual parts with a solution prepared ac-
cording to the manufacturer's instructions.
If the recommended application period is exceeded signi-
ficantly, the plastic parts may take on the smell of the me-
dium used.
RINSING:
CAUTION
Risk of allergic reactions and irritation of the mucous
membrane by disinfectants
Disinfectants can trigger allergic reactions or irritation of the
mucous membrane on contact with the skin.
• Rinse the product thoroughly to ensure that no residues of
the disinfectant remain on the PARI product.
– 35 –
®
basic