Preparation
Dismantle the inhalation device into its individual parts.
Precleaning
All individual parts must be precleaned immediately after use.
EQUIPMENT:
– Drinking water temperature of about 15 °C
PROCEDURE:
• Rinse all parts used for 2 minutes in running drinking water.
Cleaning and disinfecting
Please observe the instructions for use for the chemicals used.
Manual cleaning
EQUIPMENT
The method has been validated in Europe using:
– pH-neutral cleaning agent:
Bode Bomix
– Application time: 10 minutes
PROCEDURE:
CAUTION
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Observe the mixing proportions indicated and the spe-
cified treatment time.
• Ensure that all components are completely submerged in
the solution for the whole of the treatment time. There
must not be any air pockets or bubbles.
®
 plus (concentration: 0.1%)
– 32 –