5.5 Visual inspection and storage
Check all individual components. Replace any broken, de-
formed or seriously discoloured parts.
Storage location:
– dry
– dust-free
– protected from sources of contamination
optional: Use sterile packaging
6
TROUBLESHOOTING
Contact the manufacturer or distributor:
– in the event of faults that are not listed in this chapter.
– if the suggested procedure does not correct the fault.
Fault
No aerosol is
coming out of
the nebuliser.
– 39 –
Possible cause
The nebuliser nozzle
is blocked.
The connection
tubing is not connec-
ted properly.
The connection
tubing is leaking.
Remedy
Clean the nebuliser.
Check that the tubing con-
nectors are connected firmly
to the compressor and the
nebuliser.
Replace the connection
tubing.