| INSTRUCTIONS FOR USE
The position of the proGAV 2.0 XABO is indi-
cated by an arrow mark. The arrow indicates
the flow direction.
For localisation, the proGAV 2.0 Compass is
placed centrally on the marking of the valve that
needs to be checked. Because of the packag-
ing, the proGAV 2.0 Compass is slightly tilted
on the packaging.
Checking and/or adjusting without unpacking
retains the sterility of the product and the
integrity of the packaging.
Fig. 16: Packaging of the proGAV 2.0 XABO: marking
of the adjustable valve with an arrow
Fig. 17: Position of the proGAV 2.0 Compass on the
packaging of the proGAV 2.0 XABO
The proGAV 2.0 XABO can be safely com-
bined with our range of implantable shunt com-
ponents. We recommend using the Christoph
Miethke GmbH & Co. KG products in combina-
tion with the proGAV 2.0 XABO.
The use of shunt systems with a reservoir
provides options for the withdrawal of cere-
brospinal fluid, administration of drugs and
pressure control.
Thanks to an integrated check valve in the
CONTROL RESERVOIR and the SPRUNG
RESERVOIR,
cerebrospinal
pumped towards the valve, thus making it pos-
sible to check the distal part of the shunt sys-
tem as well as the Ventricular Catheter.
During the pump action, access to the Ventricu-
lar Catheter is closed. The use of this Reservoir
30
does not increase the opening pressure of the
shunt system. Puncturing the Reservoir should
be performed as perpendicular as possible to
the reservoir surface with a maximum cannula
diameter of 0.9 mm. A firm titanium floor pre-
vents the bottom surface from being pierced.
30 punctures are possible without any restric-
tions.
Burrhole Deflector
Because of the tight fit on the Ventricular
Catheter, the Burrhole Deflector makes it pos-
sible to choose the length of catheter penetrat-
ing into the skull prior to implantation. The Ven-
tricular Catheter is deflected at a right angle in
the burr hole (see chapter 4.5.5).
The proGAV 2.0 XABO is supplied as a shunt
system with integrated catheters (internal diam-
eter 1.2 mm, external diameter 2.5 mm).
For a new connection of catheter and connec-
tor, the catheters have to be carefully fixed with
a ligature to the valve's Titanium Connectors.
fluid
can
be
The medical device has been designed
to work accurately and reliably over long
periods of time. We guarantee that our
medical devices are free from defects and
fully functional at the time of dispatch.
Excluded from this guarantee are cases
in which the medical device has been
exchanged or rather explanted for tech-
nical or medical reasons outside of our
sphere of responsibility.
proGAV 2.0 XABO together with the entire
shunt system can safely withstand pres-
sure of up to 100 cm H O occurring during
and after surgery.
proGAV
®
2.0 XABO
®
Reservoir