4.11 Set the signal average
Set the number of saved measurements the analyzer uses to calculate an average measurement
(1–5). At the end of the measurement cycle, the analyzer saves the average measurement to the
data log. In addition, the analyzer updates the relays and analog outputs so they represent the saved
measurement. The signal average setting decreases variability in measurements.
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > SIGNAL AVERAGE.
2. Push the UP or DOWN arrow key to set the value. The default is 1 (no signal average used).
4.12 Change the analyzer or channel names
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER.
2. Select an option.
Option
EDIT ANALYZER
NAME
EDIT CHANNEL
NAME
4.13 Start or stop measurements on a channel (2- or 4-channel analyzers)
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > CONFIGURE
SEQUENCER > ACTIVATE CHANNELS.
2. Select a channel to start measurements. Unselect a channel to stop measurements. Push the
LEFT arrow to select or unselect a checkbox.
4.14 Change the channel measurement order (2- or 4-channel analyzers)
To change the order that the channels are measured, do the steps that follow:
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > CONFIGURE
SEQUENCER > SEQUENCE CHANNELS.
2. Push the UP and DOWN arrows to select a row.
Note: S1 is the first channel measured, followed by S2, S3 and S4.
3. Push the LEFT or RIGHT arrow to select a channel.
Note: Do not select channels that contain the symbol "~" (e.g., 4-~SAMPLE4). Channels that contain the
symbol "~" are not measured.
4.15 Set the date and time
Set the date and time format and the date and time that show on the measurement screen and in the
log files.
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > SET DATE/TIME.
2. Select an option.
Option
Description
DATE FORMAT Sets the date format (YYYY= year, MM=month and DD=day) and time format (12-hour or
24-hour). Default: YYYY-MM-DD 24 hours.
DATE/TIME
Sets the date and time. Use the arrow buttons to enter the date and time.
12 English
Description
Changes the name of the analyzer. Enter a unique name, such as the analyzer
location (16 characters maximum). The analyzer name shows on the measurement
screen and the data logs.
Changes the name of the selected channel. Enter a unique name, such as the
source of the sample water (10 characters maximum). The channel name(s) shows
on the measurement screen and the data logs.

Publicidad