6.6 Install the latest HART module firmware
Install the latest HART module firmware on the analyzer.
Items to collect:
• SD card (2 GB or more)
• PC with SD card slot and internet access
1. Put the SD card in the PC.
2. Download the latest HART firmware as follows:
a. Go to http://www.hach.com.
b. Search for "NA5600sc analyzer".
c. Select the "Downloads" tab. Scroll down to "Software/Firmware".
d. Click the link for the HART module firmware download. Select Open. A Hach folder shows.
3. Copy the HACH folder to the SD card.
Note: The HART module firmware is the bin file in \HACH\Firmware\HART 0_32768.
4. Remove the SD card from the PC.
5. Hold the SD card so that the label points to the right. Put the SD card in the SD card slot on the
analyzer. Refer to
6. Push menu, then select SD CARD SETUP > UPGRADE SOFTWARE > NETWORK CARD.
7. When the installation is complete, push enter to reboot the analyzer.
8. Remove the SD card from the analyzer.
9. Install the cover for the SD card slot to keep the environmental rating of the enclosure.
Figure 11
on page 28.
English 29