The suitability of the product for effective cleaning/disinfection was proven
by BOWA with a machine method (90°C, 5 minutes) using an alkaline
cleaning agent with a surface-acting additive (neodisher
The application of other (or non-equivalent) cleaning agents or disinfection
agents is made outside the area of responsibility of the manufacturer. Only
cleaning agents which have a proven effectiveness are permitted to be used
and which are comparable with the applied system (disinfector and cleaning
agent) in the available examination.
When selecting the disinfector it is to be observed
that the disinfector has a certified effectiveness (e.g. DGHM or FDA
Approval, resp. CE certification mark according to DIN EN ISO 15883),
that as far as possible a tested program for thermal disinfection (at
least 5 min at 90 °C or A
disinfection there is a danger of disinfectant residue on the
instruments),
that the program used is suitable for the instruments and contains
sufficient rinsing cycles,
that only sterile or germ-free (max. 10 germs/ml) and endotoxinfree
(max. 0.25 endotoxin units/ml) rinsing water is used,
that the air used for drying is filtered,
that the disinfector is regularly maintained and inspected.
When selecting the applied cleaning agent system it is to be observed
that this is basically suitable for cleaning instruments,
that, as long as no thermal disinfection is used, additionally a suitable
disinfection agent with certified effectiveness is used (e.g. DGHM or
FDA Approval, resp. CE certification mark) and that this is compatible
with the applied cleaning agent,
that the chemicals used are compatible with the instruments.
It is imperative that the data given by the manufacturer of the cleaning agent
and disinfection agent concerning the specified concentrations and
effectiveness times be adhered to.
Sequence:
1.
Place the instruments into the disinfector. Ensure that the cable is not
bent or squeezed.
2.
Connect cavities to rinsing tubes and/or hoses to ensure adequate
flushing of the instrument. A lockable joint mechanism (e.g. Luer Lock)
can be used for reliable connection to the washing machine.
3.
Start the program.
4.
Remove the instruments from the disinfector after the program has
finished
5.
Check and pack as far as possible immediately after removal (after
additional drying at a clean location).
BOWA-IFU-MN030-088-ARGON-HANDLES-AND-ELECTRODES-S6-ML-20230910
-value > 3000) is used (with chemical
0
en
mediclean forte).