Instructions for Use Run-on system Tropic plus en English
Instructions for Use for Run-on system
Trademarks
Tropic plus
Caleo
are trademarks owned by Dräger.
Safety information definitions
WARNING
A WARNING statement provides important information
about a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
NOTE
A NOTE provides additional information intended to avoid incon-
venience during operation.
For your safety and that of your patients
WARNING
Any use of the medical device requires full understanding
and strict observation of all sections of these Instructions
for Use. The medical device must only be used for the pur-
pose specified under "Intended use". Strictly observe all
WARNING and CAUTION statements throughout these
Instructions for Use and all statements on medical device
labels. Failure to observe these safety information state-
ments constitutes a use of the medical device that is incon-
sistent with its intended use.
WARNING
Strictly observe the Instructions for Use of the basic device
on which this medical device is used.
Intended use
Run-on system for use with the ultrasonic nebulizer Tropic plus.
4
Assembling the run-on system
E
A
1
Plug the prepared run-on system (A) into the nebulizer cham-
ber in place of the lid. Connect up the hoses and fix the run-on
system in place with the clamp.
 The height difference between the nebulizer chamber and the
container (bag/bottle) must not exceed 1000 mm.
2
Pierce the container with the conducting set (E), e.g., Intrafix
Air by B. Braun or Caleo conducting set (MX17018). Do not
kink the hose.
 The container must be ventilated maintaining sterile condi-
tions.
Functional test
When the container has been connected, check operation of the
float:
1
Fluid runs into the nebulizer chamber.
2
Feed stops automatically when the fluid has reached its level
(range between min. and max. mark).
Instructions for Use Run-on system Tropic plus
max.
min.