Change parameters
Verify parameters/Lock
adjustment
Incomplete process
Instrument reset
Start: Safe operating
state
Unlock adjustment
Change parameters
Setup - Function test
All parameters changed by the operator are automatically stored
character strings are identical. This is used to check the character
presentation.
carried over correctly. This is used to check device communication.
After this process is terminated, the safety function is again ensured.
Warning:
If the described process was not carried out completely or correctly
(e.g. due to interruption or voltage loss), the instrument remains in an
unlocked, and thus unsafe, status.
Warning:
In case of a reset to basic settings, all safety-relevant parameters will
also be reset to default. Therefore all safety-relevant parameters must
be checked or readjusted.
6.5

Setup process

Operating sequence
carried out as follows.
Unlock adjustment
Change parameters
tern.
Generally the instrument is in safe operating state before the adjust-
ment is released.
Each parameter change requires the release of the instrument
through a PIN (see chapter "Setup steps - Lock adjustment").
In delivery status, the PIN is 0000.
instructions and the Safety Manual.
When locking the adjustment, the instrument checks the data of the
measurement loop and decides on the basis of the evaluation results
if a function test is required.
6 Functional safety (SIL)
-