Titration parameters
4.6.3
The «Titration parameter» submenu is used to determine the actual parameters of the method (Fig. 72).
Fig. 72
Generally applicable titration parameters
The following parameters are valid:
Start drift
Control factor
End criteria:
Maximum titration time
Minimum titration time
Nachlaufzeit Stop delay time
Working point
Stop drift (delta)
Stop drift tolerance
4.6.3.1 Start Drift
The value of the «start drift» in µg/min must be equal or even lower in order to start a titration. If the value is
exceeded, the TitroLine
value can be entered from 0.01 to 99 µg/min (Fig. 73).
Fig. 73
®
7500 KF trace is in the conditioning mode. The standard value is 10.00 µg/min. This
123