Diagram 2: The criterion stop drift as difference to the start drift, the stop drift tolerance as stability criterion for
the drift. Only one of the two end criteria stop drift and stop drift tolerance must be met to end the measurement.
Sample identification
4.6.4
Here you determine whether a «sample identification» is used. You can set a «manual», «automatic» and
«no sample name» (Fig. 81).
Fig. 81
With manual sample identification, the sample identification is always asked for after the start of the method (see
also  3.5 Main Menu). With the automatic sample identification a root identification is defined (here sample,
see Fig. 82), which is then automatically numbered starting with 01.
Fig. 82
After a new power-up, numbering will resume with 01.