Clamp the titration vessel to the retaining clip. Fix the titration vessel in such a way that the bottom of the titration
vessel stands directly at the upper surface of the magnetic stirrerFig. 10).
Fig. 10
Insert the electrode KF 1150 and the generator electrode (TZ 1752 or TZ 1753) into the NS 7 and NS 19
openings of the titration vessel provided for this purpose. Connect the fixed cable of the KF 1150 indicator
electrode to the µA connection. The cable for the generator electrode LB 04 NN has a green (grey) and a black
plug and is connected to the sockets of the generator measuring input marked in colour (Fig. 11).
Fig. 11