4.6.3.2 Initial waiting time (Dead-Stop titration) / Extraction time (KF)
The «control factor» is a factor for the indicator control/speed. It can be set from 1 to 128.
The pre-set value is 4. 1 = slow and exact, 128 = fast and not exact (Fig. 74).
Fig. 74
4.6.3.3 Max. Titration Time
The «maximum titration time» is used with samples that generate an increased drift in the end and when it is
not possible to achieve stable end values. The max. titration time can be adjusted from 0 to 9999 s. The standard
value is pre-set to 600 s (Fig. 75)
Fig. 75
4.6.3.4 Min. Titration Time
After the «minimum titration time» has passed, the adjusted stop criteria are checked. The min. titration time
can be adjusted from 1 to 1800 seconds. The standard value is pre- set to 60 seconds. Particularly when worked
with a KF oven the minimum titration time must be higher (Fig. 76).
Fig. 76