88
Titrationstand TM 235 KF
1.3.2
Translation of the legally binding German version
In connection with the titrator TitroLine
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin: Germany, Made in Germany
Pump:
Free volume flow- air:
Delivery pressure max.:
Flow rate liquid medium:
Stirring speed:
50 ... 1000 U/min
Hoses:
PVC- hose (outer diameter 6 x 1 mm)
PTFE- hose (outer diameter 4 x 0.5 mm)
Connections
Low voltage input 12 V / on the backside of titration stand
Power supply:
Plug connection: plug for low voltage connection – phone jack-,
Positive pole at pin contact, inside contact = 2.1 mm, USA/Japan,
Power supply via titrator TitroLine
Housing:
Material:
Polypropylene, polyester coated
Dimensions:
80 x 130 x 250 mm (W x H x D), height without stand
Weight:
1.0 kg
Ambient conditions:
Climate:
Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °C
Altitude:
Device: No restrictions
Power supply: up to 5000 m
Pollution degree: Pollution degree IP 20, indoor use only.
®
7500 KF trace
Do not use the device in hazardous locations!
flow rate 2.25 l / min
1,5 bar
approx. 0,8 l / min
®
7500 KF trace
(Release: 21. February 2020)