Boston Scientific WaveWriter Alpha Instrucciones De Uso página 12

Ocultar thumbs Ver también para WaveWriter Alpha:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 74
Implantable Pulse Generator
7. Unscrew and remove the tunneling tool handle.
8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand. Pull the
Tunneling Tool shaft out through the Straw.
9. Push the Lead or Lead Extension through the Straw, then withdraw the straw.
10. Pull the proximal end(s) out of the exit point.
11. Wipe the proximal end(s) clean.
CAUTION: Do not tunnel the Splitter.
Note: If using the 2x8 Splitter and performing a permanent trial, the Splitter tails may be tunneled to the
exit site.
Implantable Pulse Generator
92469342-02
8 of 353

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Wavewriter alpha primeSc-1216Sc-1232Sc-1416Sc-1432

Tabla de contenido