Implantable Pulse Generator
Connecting the Lead, Lead Extension, Splitter, or Connector to the 32 Contact IPG
Note: Lead Extensions, Adapters, and Splitters are not full-body MRI scan eligible.
IPG ports are labeled as follows:
A
C
For convenience, connect Leads or Splitter tails to the IPG ports corresponding to their locations, superior
versus inferior or left versus right Lead placements. For example:
• Superior Leads to upper IPG ports A or B. Inferior Leads to IPG Ports C or D.
• For the Infinion™ 16, connect the Splitter tail with laser etched bands (Contacts 1-8 of the Infinion
16 Lead) to the left Ports A or C and the unmarked Splitter tail (Contacts 9-16 of the Infinion 16
Lead) to the right Ports B or D.
• For the Infinion CX Lead, connect the tail with the single marker band (Contacts 1-8 of the Infinion
CX Lead) to the left Ports A or C and the tail with two marker bands (Contacts 9-16 of the Infinion
CX Lead) to the right Ports B or D.
• For the Artisan 2x8 Surgical Lead, connect the left side to the left Ports A or C. Connect the right
side (the laser-etched tail), Contacts 9-16, to the right Ports B or D.
• For the 4x8 Surgical Lead, connect the Lead tail as follows to the 32 Contact IPG:
- Lead tail with one marker band (Contacts 1-8) to Port A.
- Lead tail with two marker bands (Contacts 9-16) to Port B.
- Lead tail with three marker bands (Contacts 17-24) to Port C.
- Lead tail with four marker bands (Contacts 25-32) to Port D.
Example: 4x8 Surgical Lead (electrodes faced down)
4
3
2
1
1
Contacts 1-8
2
Contacts 9-16
3
Contacts 17-24
4
Contacts 25-32
Implantable Pulse Generator
92469342-02
10 of 353
B
D
5
Connects to port A
6
Connects to port B
7
Connects to port C
8
Connects to port D
8
7
6
5