• Alternatively, centering the Charger within the alignment area (i.e., the area where the Charger
does not beep) will also ensure that the Charger is aligned.
4. Secure the Charger over the IPG by using either an Adhesive Pad or the Charger Belt.
• Adhesive Pad: Remove the clear liner from the pad. Apply the white side with the blue stripe to
the back of the Charger. Then remove the beige liner from the pad. Secure the Charger over
the Stimulator by pressing the adhesive to the skin over the Stimulator.
• Charger Belt: Place the Charger into the pocket on the Charger Belt so that the Power button
is visible through the mesh fabric. Secure the Charger over the Stimulator by adjusting the
Charger Belt.
Note: If you accidentally apply the pad in the wrong location, or if the Charger Belt moves out of
alignment, the Charger will start beeping again. Use a new Adhesive Pad or readjust the belt to
place the Charger back into position.
5. When the Charger emits a series of double beeps, the Stimulator is fully charged. Turn off the Charger,
remove the Charger Belt or Adhesive Pad, and return the Charger to the Base Station. Do not confuse
the end of charge signal (a series of double beeps) with the continuous beeps that indicate that the
Charger is searching for the IPG.
Note: • The end-of-a-charge signal is a distinct double beep, and the alignment indicator is a steady
continuous signal.
• The Remote Control or the Clinician Programmer will not be able to communicate with the IPG
when charging.
Refer to "IPG Battery Life" in your Information for Prescribers manual for information on IPG battery life.

IPG Battery Status

The patient Remote Control displays the IPG battery status when communicating with the IPG. Refer
to the Clinician Remote Control Directions for Use for additional information. When the Remote Control
indicates a low battery the IPG should be recharged as soon as possible.
Failure to recharge may lead to loss of stimulation in less than 24 hours. After stimulation stops,
communication with the IPG will also cease. Until a sufficient level of charge has been attained, the IPG
may not communicate with the Remote Control.
Rechargeable Stimulator System
Implantable Pulse Generator
92469342-02