Do not cover the power source body as it may result in excessive heating. The
power source operates when the plug is inserted into an outlet. To turn it off,
remove the plug from the outlet. The outlet must be easily accessible and located
near the printer. Should a faulty condition occur, remove the plug from the outlet
immediately.
The OtoRead is designed for use with 1.5v AA/UM-3/R6 Alkaline batteries.
Rechargeable batteries are not recommended for use with the OtoRead, however should you choose
to use them, the OtoRead
requires the use of 1.6v high voltage Nickel Zinc rechargeable battery types. Other types of
rechargeable batteries such as 1.2v Nickel-Metal Hydride rechargeable batteries must not be used as
they do not provide enough power to the device.
Rechargeable batteries can typically be purchased in a 4 pack with a charger. A 3-5 hour fast charger
is recommended as this type charges the batteries at a lower current, increasing the life of the
batteries. The PowerGenix ZR-PGX5HRAA-4B, 3-5-Hour Fast Charger with 4 AA 1.6v High Voltage
Rechargeable Batteries fit this description.
Be aware that using rechargeable batteries may cause the following:
1. The low battery message generated by the instrument mail fail to be triggered. In this case,
the instrument display may go blank or the instrument may fail to power on without warning. If
this occurs, replace the current batteries with newly recharged batteries.
2. Some rechargeable batteries tend to be slightly longer in length than regular alkaline AA
batteries. This may cause the battery contacts to be pushed outwards. When returning to use
standard alkaline AA batteries, it may be found that the alkaline batteries no longer make
proper contact. Ensure that the rechargeable batteries used are the same length as standard
alkaline AA batteries.
OtoRead™ – Instructions for Use - EN

Printer AC Power Source Precautions

The printer AC power source converts main voltage (120V AC or 220V AC) to
a low voltage (7V DC) for powering the printer. Use only the power source
supplied for this purpose.
Notice - This product is for indoor use only. Do not expose to water or
excessive dust.
Notice - This product is not suitable for operation in the presence of
flammable anesthetic mixture with air or with nitrous oxide.