3. Enter the following sequence:
•
PAGE, ON/OFF, RESET, ON/OFF, PAGE.
•
User is now:
•
in the Setup menu
•
at the top Menu,"calibration".
•
There are five top menus accessible by using the:
•
Scroll (PAGE, RESET) and
•
OK (ON/OFF) buttons.
•
The five top menus in the Setup menu are:
•
Calibration
•
Alarm Setup
•
Instrument Setup
•
Turn Off Instrument
•
Go Back to Measurement.
Calibration Mode
4. To enter this page, press OK by scrolling with the PAGE and
RESET buttons.
5. It is now possible to access the following sub-menus:
•
Calibration Multiple Sensors
•
5 sensor calibration (see TABLE 5-1)
•
4 sensor calibration (see TABLE 5-1)
•
Calibration Single Sensor
•
Fresh Air Setup
•
Calibration Check
•
Go Back to Previous Menu.
NOTE: The procedure is similar for all menus.
•
The following describe the 4 sensor calibration procedure.
5-3