3. Enter the following sequence:
PAGE, ON/OFF, RESET, ON/OFF, PAGE.
User is now:
in the Setup menu
at the top Menu,"calibration".
There are five top menus accessible by using the:
Scroll (PAGE, RESET) and
OK (ON/OFF) buttons.
The five top menus in the Setup menu are:
Calibration
Alarm Setup
Instrument Setup
Turn Off Instrument
Go Back to Measurement.

Calibration Mode

4. To enter this page, press OK by scrolling with the PAGE and
RESET buttons.
5. It is now possible to access the following sub-menus:
Calibration Multiple Sensors
5 sensor calibration (see TABLE 5-1)
4 sensor calibration (see TABLE 5-1)
Calibration Single Sensor
Fresh Air Setup
Calibration Check
Go Back to Previous Menu.
NOTE: The procedure is similar for all menus.
The following describe the 4 sensor calibration procedure.
5-3