MENU to return to 4 sensor calibration
•
NEXT to perform Calibration.
9. Press:
•
VALUE to see gas Value instead of gas Flag
•
ABORT to cancel Calibration.
10.Press:
•
MEASURE to return to Measure mode
•
REPEAT to start 4 sensor Calibration again
•
MENU to return to 4 sensor calibration.
•
If Calibration is successful, the instrument asks for
confirmation.
•
If Calibration is not successful, user can:
•
REPEAT or
•
ABORT.
Figure 5-4. Calibration Page
Calibration Single Sensor
1. Access with OK.
2. Press NEXT and follow a procedure similar to that previously given
under "Calibration Multiple Sensors".
NOTE: After performing FAS, the unit will ask for Span gas; the
calibration value can be adjusted in this mode.
5-6