11. Nasal speculum

1
2
3
4
5
Device function:
1) Screw to expand the speculum
2) Swivel lens, 2.5x magnification
3) Expandable speculum
4) Internal fibre optics
5) Bayonet mount
11.1. Purpose/indication
The nasal speculum described in these instructions for use is produced for illu-
mination and therefore examination of the inside of the nose.

11.1.1. Contraindication:

There may be a risk of ignition of gases if the instrument is operated in the
presence of flammable mixtures or mixtures of pharmaceuticals.
The instrument heads and battery handles must never be placed in liquids.
Only use Riester or Riester-approved accessories/consumables.
Cleaning frequency and sequence must comply with the cleaning regulations of
non-sterile products in the respective facility. Cleaning/disinfection instructions
in the instructions for use must be observed.
The product may only be used by trained personnel.

11.1.2. Intended patient population:

The device is intended for adults and children.

11.1.3. Intended operators/users:

The nasal specula are intended exclusively for use by doctors in clinics and
medical practices.

11.1.4. Required skills/operator training:

Since only doctors use the nasal speculum, they have the appropriate quali-
fications.

11.1.5. Environmental conditions:

The device is intended for use in rooms with a controlled environment. The de-
vice must not be exposed to adverse/harsh environmental conditions.

11.2. Commissioning and function

Attach the required instrument head to the mount on the upper part of the
handle so that the two recesses on the lower part of the instrument head align
with the two protruding guide pins of the battery handle. Gently press the in-
strument head onto the battery handle and turn the handle clockwise until it
stops. The head is removed by turning it anti-clockwise. Two types of operation
are possible:
a) Quick expansion
Push the set screw on the instrument head down with your thumb. In this ad-
justment, the position of the shank of the speculum cannot be changed.
b) Individual expansion
Turn the adjusting screw clockwise until you reach the desired expansion width.
The shanks close again when the screw is turned anticlockwise.
11.3. Swivel lens
On the nasal speculum there is a swivel lens with a magnification of around
2.5x, which can simply be pulled out or reinserted into the opening provided on
the nasal speculum, as required.
11.4. Technical data for the lamp
Nasal speculum XL 2.5 V 2.5 V 750 mA avg.lifespan 15 h
Nasal speculum XL 3.5 V 3.5 V 720 mA avg.lifespan 15 h
Nasal specula LED 2.5 V 2.5 V 280 mA avg.lifespan 10,000 h
Nasal specula LED 2.5 V 2.5 V 280 mA avg.lifespan 10,000 h
55