12. Tongue depressor holder

1
2
3
4
5
Device function:
1) Light guide
2) Plastic housing
3) Sliding mechanism for spatulas
4) Internal fibre optics
5) Bayonet mount
12.1. Purpose/indication
The tongue depressor holder described in these instructions for use is manu-
factured for the examination of the mouth and throat area in combination with
standard wooden and plastic depressors.

12.1.1. Contraindication:

There may be a risk of ignition of gases if the instrument is operated in the
presence of flammable mixtures or mixtures of pharmaceuticals.
The instrument heads and battery handles must never be placed in liquids.
Only use Riester or Riester-approved accessories/consumables.
Cleaning frequency and sequence must comply with the cleaning regulations of
non-sterile products in the respective facility. Cleaning/disinfection instructions
in the instructions for use must be observed.
The product may only be used by trained personnel.

12.1.2. Intended patient population:

The device is intended for adults and children.

12.1.3. Intended operators/users:

The tongue depressor holders are intended exclusively for use by doctors in
clinics and medical practices.

12.1.4. Required skills/operator training:

Since only doctors use the tongue depressor holders, they have the appropriate
qualifications.

12.1.5. Environmental conditions:

The device is intended for use in rooms with a controlled environment. The de-
vice must not be exposed to adverse/harsh environmental conditions.
12.2. Commissioning and function
Attach the required instrument head to the mount on the upper part of the
handle so that the two recesses on the lower part of the instrument head align
with the two protruding guide pins of the battery handle. Gently press the in-
strument head onto the battery handle and turn the handle clockwise until it
stops. The head is removed by turning it anti-clockwise. Insert a commercial
wooden or plastic tongue depressor into the aperture below the light opening
up to the stop. The tongue depressor is easy to remove after examination by
actuating the ejector.
12.3. Technical data for the lamp
Tongue depressor XL 2.5 V 2.5 V 750 mA avg.lifespan 15 h
Tongue depressor holder XL 3.5 V 3.5 V 720 mA avg.lifespan 15 h
Tongue depressor LED 2.5 V 2.5 V 280 mA avg.lifespan 10,000 h
Tongue depressor LED 3.5 V 3.5 V 280 mA avg.lifespan 10,000 h
1
2

13.Laryngeal mirror

Device function:
1) Mirror
2) Bracket for lamp holder
13.1. Purpose/indication
The laryngeal mirrors described in these instructions for use are manufactured
for mirroring orexamination of the mouth and throat area in combination with
a Riester lamp holder.