CAUTION!
The light intensity is at 100%, every time the battery handle is turned on. Auto-
matic safety shutdown after 180 seconds.
Explanation of the symbol on the socket handle:
CAUTION!
Observe the instructions for use!
1
2
3
4
5
6. ri-scope® L otoscope
Device function:
1) Ear specula
2) Bayonet mount
3) Swivel lens, 3x magnification
4) OP lens
5) Push button for ear speculum ejection (L3 only)
6.1. Purpose/indication
The Riester otoscope described in these operating instructions is produced for
illumination and examination of the auditory canal in combination with Riester
ear specula.
6.2. Attaching and removing ear specula
The otoscope head can take either disposable Riester ear specula (in black) or
reusable Riester ear specula (in black). The size of the ear speculum is marked
at the back of the speculum.
Otoscope L1 and L2
Turn the speculum clockwise until you feel resistance. To remove the speculum,
turn it anti-clockwise.
Otoscope L3
Place the selected speculum on the chromed metal fitting of the otoscope until
it clicks into place. To be able to remove the speculum, press the blue eject
button. The speculum will detach automatically.
6.3. Swivel magnifying lens for enlarging
The swivel lens is fixed to the device and can be swivelled 360°.
6.4. Inserting external instruments into the ear.
If you wish to insert external instruments into the ear (e.g. forceps), you have to
turn the swivel lens (approx. 3x magnification) located on the otoscope head by
180°. You can now insert the operation lens.
6.5. Pneumatic otoscopy
Pneumatic otoscopy (= an examination of the eardrum), requires a ball which
is not included in normal delivery but may be ordered separately. The hose of
the ball is put on the connection. You can now carefully introduce the necessary
amount of air into the ear canal.
6.6. Technical data for the lamp
Otoscope XL 2.5 V, 2.5 V, 750 mA, averagelifespan 15 h
Otoscope XL 3.5 V, 3.5 V, 720 mA, averagelifespan 15 h
Otoscope LED 2.5 V, 2.5 V, 280 mA, averagelifespan 10,000 h
Otoscope LED 3.5 V, 2.5 V, 280 mA, averagelifespan 10,000 h
7. ri-scope® L ophthalmoscopes
Device function:
1
2
3
4
5
6
1) Dioptre display
2) Dioptre wheel
3) Aperture symbols
4) Aperture wheel
5) Filter wheel
6) Bayonet head
50