15 Walking Calibration:
Click on:
Calibrate
to select the Walking Calibration screen.
Walking Calibration procedure:
Move the slider to select the amputation level,
Transfemoral TF or Transtibial TT
B
1 Start Calibration (stage 1)
To calibrate the Elan to the user's
individual gait characteristics instruct
the user to walk on a horizontal and level
surface and at a comfortable pace.
After several steps,
click on:
Press to start calibration
C
The step counter counts down the
number of remaining steps required.
(Note: 8-10 uninterrupted, regular steps
are required for calibration).
Continue to Stage 2
B
D
earlier versions
D
30
Calibration may be repeated
at any time to fine tune the
set-up and is recommended
after any change in
alignment and/or basic
resistance settings.
C
The Elan software adapts
to the individual user's gait
characteristics from the
Calibration stages.
(See Section 3 for more
information)