Transferring the Mattress from one Bed to
Another
Never leave the patient on the mattress during transfer.
Make sure that the bed chassis brakes are applied to prevent
any accidental movements.
If the mattress has handles, use them to transfer the mattress
more easily.
While handling the mattress, all necessary precautions must
be taken to avoid any damage.
Do not work on the device when it is occupied.

Cleaning and Disinfecting

Cleaning and disinfection are two separate procedures.
Complying with Safety Instructions
Check that the brakes are applied on the bed on which the
mattress is installed.
Lock out all electric functions.
Never clean the mattress by pouring water on it, nor with a
high-pressure jet.
Never use water at a temperature of more than 70°C.
Refer to the recommendations of the cleaning and
disinfecting of the product provided by the manufacturer.
Thoroughly dry the mattress before reusing it.
Suitable individual protective equipment must be worn
during the steps of the cleaning operation (blouse, gloves,
eye protection, etc.).
Failure to implement one or more of these
recommendations may lead to damage or deterioration,
preventing use of the device and rendering the warranty
void.
Controlling Infection
Insufficient Cleaning = Risk of Infection
(biological danger)!
All parts must be kept clean at all times in order to avoid the
risk of infection. All necessary precautions must be taken to
eliminate all visible soiling.
The following recommendations are not designed to
replace cleaning and disinfection protocols drawn up by
the hygien officer or by other bodies for your hospital in
the event of particular infectious situations.
Complying with Manufacturer Recommendations
The cleaning and disinfecting method described below applies
specifically to the mattress and its accessories and is
designed to save time and to help combat nosocomial
infection more effectively.
Hill-Rom recommends that the mattress be disinfected
before first use.
When cleaning, always check the cover for cuts, tears,
cracks or snags; never use a mattress with a damaged
cover.
If the internal foam of the mattress is soiled and presents
a risk of infection, then replace the mattress.
Following other cleaning and disinfection protocols or
using other products than those recommended by
Hill-Rom may compromise devices and patient safety,
and void the warranty.
© 2021 Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
April 2021
De-commissioning
Hill-Rom recommends that the mattress should be disinfected
before de-commissioning.
Customers should adhere to all federal, state, regional, and/or
local laws and regulations as it pertains to the safe disposal of
medical devices and accessories. If in doubt, the user of the
device shall first contact Hill-Rom Technical Support for
guidance on safe disposal protocols.
Products To Be Avoided
Never use cleaning agents, detergents, degreasing agents or
industrial solvents containing any of the following products in
order to avoid damaging the mattress:
Phenol
Hydrochloric,
Nitric or Sulfuric
Acids
Dimethylformamide
Do not use highly acidic detergents or disinfectants (pH<4).
Do not use highly basic detergents or disinfectants (pH>10).
Never use abrasive cleaning materials or products, such as
scouring pads.
Cleaning and Disinfecting after the Departure of a
Patient or when Changing the Sheets
Using Recommended Products
Terralin® Protect (Schülke)
Concentration: 5 ml/l (= 0.5%), contact time: 1 h.
Cleaning
Clean the mattress with a cloth slightly dampened
with hot water and a neutral detergent solution.
Check that the solution does not contain any of
the products to be avoided mentioned above.
Cleaning Tough Stains
Quickly wipe away any traces of pharmaceutical solutions
used for the patients to avoid permanent damage to the
surface.
Remove tough stains with a soft brush and neutral
detergents or a chlorine-based solution with a
concentration less than or equal to 1,000 ppm or
10,000 ppm according to labeling.
To eliminate hardened stains (excreta, other forms of
soiling), soften them by soaking and take care to
thoroughly dry the cover before putting it back on the
mattress.
Disinfecting
In the event of visible soiling, Hill-Rom recommends that the
mattress should be disinfected with an intermediate
disinfectant (tuberculocidal) that complies with the applicable
regulations (e.g. with the requirements of regulation
2017/745/EEC).
For all other disinfectants:
Proceed with the final steps before reusing the cleaned and
disinfected mattress (See "Performing the Final Steps",
page 3).
1. Terralin® Protect is listed by the V.A.H (Verbund für
Angewandte Hygiene e.V.).
163791(10) - EN
Cresol
Soda
Tetrahydrofuran
1
Chlorine-based solutions can be used.
Concentration must be less than or equal to
1,000 ppm.
Ethanol (alcohol) can be used.
Concentration must not exceed ¼ of ethanol
for ¾ of water.
Page 2 - 4