Fig. 8:
Position adsorber
1 Diagnostic module
1 Frame
1 Adsorber
Fig. 9:
Adsorber
1 Drying agent
2 Drying agent label
i
light grey drying agent: replace adsorber.
4. Unscrew used adsorber.
5. Unscrew protective cap of the new drying agent
adsorber.
Fig. 10:
Unscrewing the protective cap
1 Protective cap
2 Adsorber
Robert Bosch GmbH
Maintenance | TTM 2104 / TTM 2204 | 45 | en
6. Screw new adsorber in threaded hole of the
measuring unit.
7. Attach new drying agent label to the top side of the
adsorber.
8. Remove adsorber's red protective backing.
!
Do not remove protective backing until
commissioning.
9. Close cover plates.
10. Hinge drive-down ramps in measuring units.
7.2.5
Replacing slide protection strips
1. Remove displaced or worn slide protection strips
from the cover plate.
2. Remove protective strip of the new slip protection
strip.
3. Attach new slide protection strip to cover plate.
Fig. 11:
Replacing slide protection strips
1 Slide protection strip
2 Cover plate
7.2.6
Activating the measurement mode
1. Select <Settings>.
2. Select operation mode Measurement in General
information tab.
3. Select <Apply>.
|
1 691 206 005
2019-02-25